Measurement Ranges
Element | Range | Typical Detection Limit | Suitable For |
Carbon (C) | 0.1 ppm to 6% | ~0.1 ppm / ±0.5 ppm | From ultra-low impurity to high-alloy steels |
Sulfur (S) | 0.1 ppm to 30% | ~0.1 ppm / ±0.8 ppm | Steel, cast iron, ores, refractories |
Ranges are based on ~500 mg sample weight; adaptable to different sample sizes.
Analysis Speed
50 seconds per sample (cycle ~60 seconds)
Optimized for production floor feedback loops and high-throughput labs.
Core Technology
Component | Function | Notes |
Induction Furnace (≈2200°C) | Complete combustion of sample | PC-controlled heating, up to 2.2 kVA power NCS-Combustion-Master-CS-EN |
Pure Oxygen Flow | Ensures full oxidation to CO₂ & SO₂ | Delivered through precision lance |
Gas Conditioning System | Dust and moisture removal | Supports stable detector performance NCS-Combustion-Master-CS-EN |
IR Detector Units | Measures CO₂ and SO₂ concentrations | Up to 4 independent detectors for wide-range flexibility NCS-Combustion-Master-CS-EN |
The analyzer calculates C & S content directly from IR absorption data, ensuring high accuracy and reproducibility.
Software & Automation
PC-controlled workflow
Automatic furnace cleaning
Automatic leak testing built-in
Customizable methods, user logins, QC modes
Optional remote diagnostic service support
Applications
Sector | Usage |
Steel & Iron Production | Process monitoring from blast furnace → converter → caster NCS-Combustion-Master-CS-EN |
Foundries | Rapid verification of molten metal chemistry |
Mining / Geology | Ore grading and beneficiation control NCS-Combustion-Master-CS-EN |
Catalyst & Chemical Industry | Elemental impurity screening |
| Component | Function | Notes |
|---|---|---|
| Induction Furnace (≈2200°C) | Complete combustion of sample | PC-controlled heating, up to 2.2 kVA power |
| Pure Oxygen Flow | Ensures full oxidation to CO₂ & SO₂ | Delivered through precision lance |
| Gas Conditioning | Dust and moisture removal | Supports stable detector performance |
| IR Detector Units | Measures CO₂ and SO₂ concentrations | Up to 4 independent detectors for wide-range flexibility NCS-Combustion-Master-CS-EN |
Operational Advantages
Feature | Why It Matters |
Fast analysis cycle | Real-time process adjustment in steelmaking |
High thermal and signal stability | Repeatable, traceable results |
Modular detector configuration | Scale according to application depth |
Rugged industrial build | Designed for continuous 24/7 metallurgical workflow |
Simple maintenance access | Reduced downtime in production environments |
Standards Compliance
Supports major international methods including:
ASTM E1019, E1587, E1941, E1915, UOP 703
ISO 9556, 10719, 15349-2, 4935, 13902, 10694, 4689-3